Model Number TABLETOP-JAPAN |
Device Problems
Backflow (1064); Decrease in Pressure (1490)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/10/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that during surgery an ophthalmic surgical console presented with unstable intraocular pressure control and air was flowed backward.The procedure was completed and patient harm was not reported.
|
|
Manufacturer Narrative
|
The company representative did not confirm nor replicate anything that would have contributed to the reported event.The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.The system was found to meet specifications.Therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|