SMITH & NEPHEW, INC. R3 0 HOLE HA CTD ACET SHELL 46MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
|
Back to Search Results |
|
Catalog Number 71332246 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Bone Fracture(s) (1870)
|
Event Date 03/15/2014 |
Event Type
Injury
|
Event Description
|
It was reported that, after a left thr surgery performed on (b)(6) 2014 to address osteoarthritis symptoms, the patient sustained a peri-prosthetic fracture of the socket that made necessary a revision surgery on (b)(6) 2014.During this procedure, the modular head, acetabular liner and acetabular cup were explanted.This information was provided by the national joint registry of the united kingdom, as part of a retrospective data collection of patients who underwent a primary thr surgery with a hip prosthesis construct that included a polarstem stem with an r3 acetabular cup and that required a revision surgery due to specific reasons.As such, no further information will be available.
|
|
Manufacturer Narrative
|
This complaint was opened by smith+nephew to document a patient complication identified through a review of the national joint registry uk from united kingdom that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.(b)(4).
|
|
Search Alerts/Recalls
|
|
|