• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL Back to Search Results
Model Number M00510860
Device Problems Separation Failure (2547); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a trapezoid rx basket was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, a trapezoid basket was used to crush a stone.However, the basket was difficult to open and close when the handle was actuated.The stone was eventually freed from the basket after several attempts.The basket was removed, and the procedure was completed with another trapezoid basket.There were no patient complications reported as a result of this event.On (b)(6) 2023, it was confirmed that the tip of the basket did not detach when an alliance handle was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16141244
MDR Text Key307901205
Report Number3005099803-2023-00067
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296379
UDI-Public08714729296379
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/06/2023
Device Model NumberM00510860
Device Catalogue Number1086
Device Lot Number0030088091
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/09/2023
Initial Date FDA Received01/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/06/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-