The reporter indicated the surgeon implanted a 12.6mm vticm5_12.6 implantable collamer lens, -14.00/+1.0/090 (sphere/cylinder/axis), into the patients right eye (od) on (b)(6) 2022.The lens was removed on (b)(6) 2022 due to excessive vaulting, elevated intraocular pressure (iop), pupil block, angle closure, unreactive (fixed) pupil and iris atrophy.An additional pi was performed on (b)(6) 2022 and there was anterior chamber irrigation/evacuation of visco/fluids.The lens was replaced with a shorter lens but the problem was not resolved.See mfr.Report # 2023826-2023-00099 for replacement lens.Remaining pupil dissipation after an acute glaucoma attack.Additional information has been requested but none has been forthcoming.
|
H3: device evaluation - lens was returned in microcentrifuge vial dry.Visual inspection found haptic broken.Unable to perform dimensional inspection due to signficant lens haptic damage.H6: device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device; have been manufactured within established process parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim# (b)(4).
|