• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Degraded (1153)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2022
Event Type  malfunction  
Manufacturer Narrative
The device was evaluated and repaired on-site by an olympus field service engineer; the grey connector was replaced.The device is not expected to be returned.The investigation is ongoing and follow-up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus that the grey connector fell apart on the endoscope reprocessor.There was no patient harm associated with the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The cause of the reported event cannot be conclusively determined.Based on the results of the investigation, it is likely the connector could not be connected as some impact was added to the connector and the connector became loose and came apart.Probable causes for the event include: "* accumulated stress over time which caused the connector to get loose * unintentional impact to the connector with something hard." per the instructions for use: ¿check the following for each connector.The connector should be fixed firmly.The o-rings should be free of abnormalities such as cracks, tears, or dents.¿ and ¿warning -do not use the equipment if any connector seems to be damaged or defective.Using the equipment when an irregularity has been detected may interfere with reprocessing.Furthermore, fluid leakage may damage peripheral device or facilities near the equipment.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16145631
MDR Text Key307164086
Report Number9610595-2023-00656
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/15/2022
Initial Date FDA Received01/12/2023
Supplement Dates Manufacturer Received01/19/2023
Supplement Dates FDA Received02/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/27/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-