• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLFLEX DUODENAL; STENT,METALLIC,EXPANDABLE,DUODENAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLFLEX DUODENAL; STENT,METALLIC,EXPANDABLE,DUODENAL Back to Search Results
Model Number M00565030
Device Problems Break (1069); Activation, Positioning or Separation Problem (2906); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2022, that a wallflex enteral duodenal stent was to be implanted in the duodenum to treat a 3 cm extrinsic malignant duodenal stricture due to pancreatic cancer during an esophagogastroduodenoscopy (egd) with stent placement procedure performed on (b)(6) 2022.The patient's anatomy was tortuous and was not dilated prior to stent placement.During the procedure, the delivery handle snapped in two places, and the control hub was noted to be broken.The wallflex duodenal stent was unable to fully deploy and was removed from the patient partially covered by the outer sheath.The procedure was cancelled, and the patient was scheduled for another stent placement at another healthcare facility.There were no patient complications reported as a result of this event.Note: a photo of the complaint device was provided by the complainant and showed the stainless-steel handle was slightly bent, the blue outer sheath was separated from the exterior tube handle, and the wallflex duodenal stent was partially deployed outside the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX DUODENAL
Type of Device
STENT,METALLIC,EXPANDABLE,DUODENAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16151972
MDR Text Key308014796
Report Number3005099803-2022-08074
Device Sequence Number1
Product Code MUM
UDI-Device Identifier08714729456506
UDI-Public08714729456506
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062750
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00565030
Device Catalogue Number6503
Device Lot Number0030457353
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/21/2022
Initial Date FDA Received01/12/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-