BAXTER HEALTHCARE CORPORATION VALVE SET, EM2400; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938724
Device Problems Protective Measures Problem (3015); Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2022
Event Type  malfunction  
Event Description
It was reported that there were air bubbles detected in two (2) em2400 valve sets.During setup, air bubbles were observed in the valve set.Tech support assisted the customer with replacing the set with a new one of the same lot; however, air bubbles were detected again.To resolve this issue, the set was replaced with a new one from a different lot and the customer was able to continue with setup.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information was added to h4, h6 and h10.Correction to b5 and f10/h6: medical device problem codes (update a1415 to a16).B5: upon further review, the valve set triggered bubble detected alarms on the automated compounding device; there was no mention of bubbles in the set.H4: device manufacture date: august 31, 2022 - september 2, 2022.H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The devices were not received for evaluation; therefore, a device analysis could not be completed.However, this issue did not result in a leak, contamination or accuracy issue and is not likely to cause or contribute to a death or serious injury.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALVE SET, EM2400
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16152034
MDR Text Key307721057
Report Number1416980-2022-07377
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477183
UDI-Public(01)00085412477183
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K002705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938724
Device Lot Number60392782
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/16/2022
Initial Date FDA Received01/12/2023
Supplement Dates Manufacturer Received02/02/2023
Supplement Dates FDA Received02/09/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-