Model Number 39467-175
Device Problems Break (1069); Entrapment of Device (1212)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2022
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.
 
Event Description
It was reported that the burr got stuck on lesion.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified proximal right coronary artery (rca) to atrioventricular branch.A 1.75mm rotapro was selected for use.During the procedure, the device got stuck in the lesion while cutting the proximal rca and the non-boston scientific guiding catheter was pushed to release the device that got stuck.During the process it was also found that the driveshaft was disconnected, therefore, the entire system was removed.The procedure was completed with another of the same device.No complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAPRO
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16152875
MDR Text Key308359688
Report Number2124215-2022-55668
Device Sequence Number1
Product Code MCX
UDI-Device Identifier08714729893356
UDI-Public8714729893356
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number39467-175
Device Catalogue Number39467-175
Device Lot Number0029672233
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/20/2022
Initial Date FDA Received01/12/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GUIDE CATHETER-MEDTRONIC LAUNCHER; GUIDEWIRE-BSJ ROTAWIRE DRIVE; IMAGING CATHETER-BSJ OCHD US
Patient SexFemale
-
-