Catalog Number 2C2063K
Device Problems Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2022
Event Type  malfunction  
Event Description
It was reported that the fill port of a large volume infusor was damaged and resulted in a fluid leak.This issue was further described as, ¿they noticed there was a leak at the connection point it looks like there is a split in the hard plastic connector¿.This event occurred while compounding the device.The device was filled with sodium chloride 180ml.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Initial reporter postal code: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Correction: g4: pma/510k #.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection was performed to the photograph which showed white drug residue around the fill port area which suggests a leak may have occurred.Additionally, the photo showed cracks on the fill port which resulted in leak.The reported condition was verified.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16157126
MDR Text Key308580357
Report Number1416980-2022-07379
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C2063K
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/21/2022
Initial Date FDA Received01/13/2023
Supplement Dates Manufacturer Received02/06/2023
Supplement Dates FDA Received02/17/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-