Model Number 8886848813
Device Problem Device Slipped (1584)
Patient Problems Abdominal Pain (1685); Peritonitis (2252); Unspecified Tissue Injury (4559); Unintended Radiation Exposure (4565)
Event Date 12/05/2022
Event Type  Injury  
Event Description
According to the reporter, one day postoperative to laparoscopic cholecystectomy, the patient developed upper abdominal pain and was given symptomatic treatment.On the second day, the abdominal pain worsened and computed tomopgraphy (ct) and paracentesis were performed, and bile leakage and peritonitis were found.Exploratory laparotomy revealed a large amount of bile-like fluid in the abdominal cavity.The cystic duct entered the common bile duct, the absorbable ligature clip fell off, and bile leaked into the abdominal cavity.Peritoneal washing and drainage were performed.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPRO-CLIP
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16163469
MDR Text Key307465265
Report Number1219930-2023-00226
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10884521057661
UDI-Public10884521057661
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K925602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8886848813
Device Catalogue Number8886848813
Device Lot NumberN1G0494Y
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/29/2022
Initial Date FDA Received01/13/2023
Date Device Manufactured07/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN LAPRO-CLIP INSTRUMENT (LOT#UNKNOWN)
Patient Outcome(s) Required Intervention;
Patient Age37 YR
Patient SexFemale
-
-