ALPHATEC SPINE, INC. INVICTUS; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 15100
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Failure of Implant (1924)
Event Date 11/09/2022
Event Type  Injury  
Event Description
At 6-months postoperatively, a radiograph revealed a set screw had backed out of an iliac bolt.A revision surgery was performed on (b)(6) 2022 to remove and replace the set screw.
 
Manufacturer Narrative
The implant was used for treatment, not diagnosis.It did not return to alphatec for evaluation.Radiographs were provided which confirmed the event.The identifying lot number was not provided; therefore, a review of the device history records could not be conducted.Based on the information provided, a root cause could not be determined.If additional information is provided, a supplemental report will be submitted.Labeling review: "intraoperativemanagement: final tightening of set screws: all set screws must be tightened using the appropriate instruments (e.G., torque handle, final driver, and counter torque) as indicated in the surgical technique guide.Warnings/cautions/precautions: risks identified with the use of these devices, which may require additional surgery, include device component failure, loss of fixation/stabilization, non-union, vertebral fracture, neurological injury, vascular or visceral injury.Do not final tighten under compression or distraction as the rod may not be normalized to the tulips, resulting in rod slippage.Failure to tighten set screws using the recommended instrument(s) could compromise the mechanical stability of the construct.Possible adverse effects: the following complications and adverse reactions have been shown to occur with the use of similar spinal instrumentation.These effects and any other known by the surgeon must be discussed with the patient preoperatively.1.Initial or delayed loosening, disassembly, bending, dislocation, and/or breakage of device components.]".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVICTUS
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ALPHATEC SPINE, INC.
1950 camino vide roble
carlsbad CA 92008
Manufacturer (Section G)
ALPHATEC SPINE, INC.
1950 camino vide roble
carlsbad CA 92008
Manufacturer Contact
wesley channell
1950 camino vide roble
carlsbad, CA 92008
9014283693
MDR Report Key16164288
MDR Text Key307464835
Report Number2027467-2022-00107
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00190376137902
UDI-Public(01)00190376137902
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181677
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number15100
Device Catalogue Number15100
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/16/2022
Initial Date FDA Received01/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-