BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number L311
Device Problems Over-Sensing (1438); Pacing Problem (1439); Incorrect, Inadequate or Imprecise Result or Readings (1535); Inappropriate or Unexpected Reset (2959); Interrogation Problem (4017)
Patient Problems Loss of consciousness (2418); Syncope/Fainting (4411)
Event Date 01/08/2023
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient had presented to the hospital due to loss of consciousness.The pacemaker was unable to interrogate with a consult device or a programmer.It was later determined that this pacemaker had entered safety mode, and it was suspected that the patient's syncopal episode was secondary to unipolar oversensing with pacing inhibition.Additional information received reported that this pacemaker was explanted and replaced.No additional adverse patient effects were reported.This pacemaker is not expected for device return.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16174922
MDR Text Key307467085
Report Number2124215-2023-01503
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526559228
UDI-Public00802526559228
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial
Report Date 01/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/16/2018
Device Model NumberL311
Device Catalogue NumberL311
Device Lot Number231353
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/09/2023
Initial Date FDA Received01/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/30/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age78 YR
Patient SexMale
-
-