Model Number 77008
Device Problems Crack (1135); Increase in Pressure (1491)
Patient Problems Hemorrhage/Bleeding (1888); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/02/2023
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during use when initiating cardiopulmonary bypass, this dlp pediatric one-piece arterial cannula was reported to have feedback pressures that were too high and the customer observed that the aortic cannula was partially severed at its lower part by the thread of suture used to hold it to the skin. it was stated that this required an emergency aortic cannula change during the surgery.There was no patient impact reported with this event.It was stated that the patient was treated for ventricular septal closure.Medtronic received additional information that there was no damage to the packaging (outer box, inner packaging or sterile barrier).The cannula was not heated or cooled prior to use.
 
Manufacturer Narrative
There was a small amount of patient blood loss because of the issue.The patient did not require a transfusion because of this issue.No air entered the circuit because of the issue with the cannula.Conclusion: medtronic cannot confirm or deny the complaint of the cannula body being severed by the suture thread as no photo has been provided and no product has been returned to date.A root cause of this occurrence cannot be determined without returned product, however, from the details provided the cause of this occurrence is most likely a result of suturing on the cannula body.Review of the device history record found no abnormalities during manufacturing that would cause or contribute to the reported event.Complaints received for similar model numbers were reviewed and showed no trends warranting escalation related to this occurrence.Medtronic has made its supplier aware of this occurrence and will continue to monitor for future occurrences and trends.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PED. ARTERIAL CANNULA
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16182747
MDR Text Key307485468
Report Number2184009-2023-00041
Device Sequence Number1
Product Code DWF
UDI-Device Identifier00763000091101
UDI-Public00763000091101
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K024069
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number77008
Device Catalogue Number77008
Device Lot Number2022020174
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/10/2023
Initial Date FDA Received01/17/2023
Supplement Dates Manufacturer Received01/25/2023
Supplement Dates FDA Received04/06/2023
Date Device Manufactured02/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-