Model Number CAR-172-C
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hypersensitivity/Allergic reaction (1907); Low Blood Pressure/ Hypotension (1914); Low Oxygen Saturation (2477)
Event Date 12/29/2022
Event Type  Injury  
Event Description
A report was received on 02 jan 2023 from a healthcare professional regarding a 76 year old female patient with multiple comorbidities, who stated the patient experienced a suspected allergic reaction on (b)(6) 2022.Additional information was received from the healthcare professional on 10 jan 2023 stating the patient experienced decreased blood pressure (165/60mmhg to 77/43mmhg) and decreased oxygen saturation levels (99% to 87%) five minutes into their first treatment using the nxstage device.Treatment was terminated, 100mg of intravenous (iv) hydrocortisone was administered to the patient, and symptoms resolved.The patient changed to a dialyzer from a different manufacturer and continues to treat with the nxstage system.
 
Manufacturer Narrative
A review of the device history record (dhr) was conducted which confirmed that the product met all quality criteria and manufacturing specifications prior to release.There is no information to indicate that a malfunction occurred.Allergic or adverse reactions are known risks of hemodialysis.The nxstage user guide and instructions for use include allergic reaction as a potential risk associated with dialysis therapy and also include warnings to monitor for potential allergic reactions.Biocompatibility of the device has been established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
9784505276
MDR Report Key16187796
MDR Text Key307521466
Report Number3003464075-2023-00008
Device Sequence Number1
Product Code KDI
UDI-Device IdentifierM535CAR172C0
UDI-Public+M535CAR172C0/$$052311077052/
Combination Product (y/n)N
Reporter Country CodeSA
PMA/PMN Number
K140526
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/01/2023
Device Model NumberCAR-172-C
Device Catalogue NumberCAR-172-C
Device Lot Number11077052
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age76 YR
Patient SexFemale
-
-