• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 1716000J
Device Problems Loss of or Failure to Bond (1068); Fracture (1260); Material Separation (1562); Separation Failure (2547); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2022
Event Type  malfunction  
Manufacturer Narrative
The catalog number has not been cleared in the us, but is similar to the powerport slim implantable port, chronoflex single-lumen, kit, 6f products that are cleared in the us.The pro code and 510k number for the powerport slim implantable port, chronoflex single-lumen, kit, 6f products is identified.As the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.Device pending return.
 
Event Description
It was reported that during a port placement procedure, the valve of the sheath allegedly damaged upon peeling.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport slim implantable port, chronoflex single-lumen, kit, 6f products that are cleared in the us.The pro code and 510k number for the powerport slim implantable port, chronoflex single-lumen, kit, 6f products is identified in d2 and g4.H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one fully peeled 6.5f introducer peel-apart sheath and one vessel dilator was returned for evaluation.Gross visual evaluation was performed.Bunching and material deformation was noted throughout both peel-apart sheaths.The investigation is confirmed for the reported sheath broken and identified separation failure issue as the valve was noted to be detached from the t-handle.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: g3, h6 (device) h11: h6 (method, result, conclusion) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a port placement procedure, the valve of the sheath allegedly damaged upon peeling.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport slim implantable port, chronoflex single-lumen, kit, 6f products that are cleared in the us.The pro code and 510k number for the powerport slim implantable port, chronoflex single-lumen, kit, 6f products is identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.H10: d4 (expiry date: 06/2023).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a port placement procedure, the valve of the sheath allegedly damaged upon peeling.There was no reported patient injury.
 
Event Description
It was reported that during a port placement procedure, the valve of the sheath allegedly damaged upon peeling.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport slim implantable port, chronoflex single-lumen, kit, 6f products that are cleared in the us.The pro code and 510k number for the powerport slim implantable port, chronoflex single-lumen, kit, 6f products is identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one fully peeled 6.5f introducer peel-apart sheath and one vessel dilator was returned for evaluation.Gross visual evaluation was performed.Bunching and material deformation was noted throughout both peel-apart sheaths.Manufacturing site evaluation of the sample found that the top cap was detached from one side of the handle, where the fr of the sleeve is indicated, it was observed that after the top cap was detached, the valve was not present in the sample, traces of ultrasonic welding were observed in the middle of the detached top cap in the section that joins the handle.Therefore the investigation is confirmed for the identified material separation and loss or failure to bond issue as the valve was noted to be detached from the t-handle.However the investigation is inconclusive for the reported material integrity problem as there was no objective evidence obtained during investigation.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: (expiry date: 06/2023).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16191504
MDR Text Key307571534
Report Number3006260740-2022-06078
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741111631
UDI-Public(01)00801741111631
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K072549
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 08/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1716000J
Device Lot NumberREGQ2806
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/22/2022
Initial Date FDA Received01/18/2023
Supplement Dates Manufacturer Received03/14/2023
03/29/2023
08/29/2023
Supplement Dates FDA Received03/26/2023
04/21/2023
08/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-