This device is classified as import for export, therefore 510k is not applicable.Model eb-1570k is available in the usa with a 510k number k131028.We checked the returned unit and confirmed that the ccd module as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module.In addition, we confirmed that the distal body broken, the light guide fiber bundle (lcb) broken, the objective lens unit broken, the u/d lock lever broken, the insertion flexible tube (ift) buckled, the remote control buttons cut, and the light guide fiber bundle (lcb) loosened; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|