BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C2063K
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the tubing of a large volume infusor broke off.This issue occurred when staff was straightening the tubing during set up.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H4: the lot was manufactured from january 03, 2022 - january 05, 2022.H10: the device was received for evaluation.Visual inspection was performed using the naked eye and the tubing was observed cut off in two pieces.The cut was observed to be jagged, not smooth.The reported condition was verified.The cause of the condition could not be determined; however, the most probable cause was due to how the product was being handled upon removing from the package.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16195732
MDR Text Key307633048
Report Number1416980-2023-00069
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412579412
UDI-Public(01)00085412579412
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 02/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C2063K
Device Lot Number22A001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/23/2022
Initial Date FDA Received01/18/2023
Supplement Dates Manufacturer Received02/10/2023
Supplement Dates FDA Received02/17/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-