LIVANOVA USA, INC. VAGUS NERVE STIMULATOR BATTERY; STIMULATOR, AUTONOMIC NERVE, IMPLANTED FOR EPILEPSY Back to Search Results
Device Problems Loose or Intermittent Connection (1371); Firing Problem (4011)
Patient Problem Failure of Implant (1924)
Event Date 01/05/2023
Event Type  malfunction  
Event Description
The patient reported that she no longer had sensation that the vagal nerve stimulator was firing.On [dated redacted], the patient was taken to the or for revision of the stimulator.Findings during the surgical procedure: functional lead with a loose connection to the generator.Due to a possible malfunction of the locking screw, it was replaced, and a new model 1000 generator and torque wrench was used to tighten and locked the lead.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAGUS NERVE STIMULATOR BATTERY
Type of Device
STIMULATOR, AUTONOMIC NERVE, IMPLANTED FOR EPILEPSY
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
MDR Report Key16203653
MDR Text Key307677455
Report Number16203653
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/06/2023
Event Location Hospital
Date Report to Manufacturer01/19/2023
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/19/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age14965 DA
Patient SexFemale
Patient Weight85 KG
Patient RaceWhite
-
-