OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, PK TURIS; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE-USE
|
Back to Search Results |
|
Model Number WA22606S |
Device Problem
Failure to Deliver (2338)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/23/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is not expected to be returned for evaluation.The investigation is ongoing and follow-up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Event Description
|
The customer reported to olympus that while using the hf-resection electrode "plasmaloop", (loop, medium, 24 fr., 12°-30°, pk turis), the cautery piece was not cauterizing well.The issue occurred during the diagnostic procedure (transurethral bladder tumor excision), and the procedure was completed with a similar device.There was no reported patient harm associated with the event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the cause of the reported issue is attributable to wear and tear.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
|
|