Model Number 00-1026-R |
Device Problems
Insufficient Information (3190); Device Handling Problem (3265)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/22/2022 |
Event Type
malfunction
|
Event Description
|
During preventive maintenance activity, the engineer observed ibp was not detected in the device.There was no patient involved as it was observed during pm activity.The complaint is still under investigation.A final report will be submitted once the investigation is complete.
|
|
Event Description
|
During preventive maintenance activity, the engineer observed ibp was not detected in the device.There was no patient involved as it was observed during pm activity.The complaint is still under investigation.A final report will be submitted once the investigation is complete.
|
|
Manufacturer Narrative
|
This report is based on information provided by philips repair service quest international personnel and has been investigated by the philips complaint handling team.Philips received a complaint on the tempus pro indicating that ibp is no longer recognized.Ibp was enabled on this device.Problem detected during the preventive maintenance.Furthermore, technical investigation found out that the tactical switch was enabled causing ibp will not working.When disabling the tactical switch it works as expected.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The device was operational after applying the change.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
|