• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: REMOTE DIAGNOSTIC TECHNOLOGIES LTD. TEMPUS PRO; MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

REMOTE DIAGNOSTIC TECHNOLOGIES LTD. TEMPUS PRO; MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number 00-1026-R
Device Problems Insufficient Information (3190); Device Handling Problem (3265)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/22/2022
Event Type  malfunction  
Event Description
During preventive maintenance activity, the engineer observed ibp was not detected in the device.There was no patient involved as it was observed during pm activity.The complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Event Description
During preventive maintenance activity, the engineer observed ibp was not detected in the device.There was no patient involved as it was observed during pm activity.The complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Manufacturer Narrative
This report is based on information provided by philips repair service quest international personnel and has been investigated by the philips complaint handling team.Philips received a complaint on the tempus pro indicating that ibp is no longer recognized.Ibp was enabled on this device.Problem detected during the preventive maintenance.Furthermore, technical investigation found out that the tactical switch was enabled causing ibp will not working.When disabling the tactical switch it works as expected.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The device was operational after applying the change.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS PRO
Type of Device
MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK  GU14 6XW
Manufacturer (Section G)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK   GU14 6XW
Manufacturer Contact
rekha rv
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key16216319
MDR Text Key307837136
Report Number3003832357-2023-00017
Device Sequence Number1
Product Code MHX
UDI-Device Identifier05060472441058
UDI-Public05060472441058
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K201746
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00-1026-R
Device Catalogue Number989706000101
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/22/2022
Initial Date FDA Received01/20/2023
Supplement Dates Manufacturer Received12/22/2022
Supplement Dates FDA Received02/23/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-