• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Catalog Number AB2000
Device Problems Connection Problem (2900); Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2022
Event Type  malfunction  
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation (procept) became aware that during the aquablation procedure, the aquabeam handpiece was unable to be engaged to the aquabeam motorpack despite multiple attempts being made.The handpiece was replaced with a new handpiece unit and the procedure was continued through successful completion.The reported event caused a surgical procedural delay of over 20 minutes.There were no adverse health consequences to the patient due to the procedural delay.
 
Manufacturer Narrative
Additional manufacturer narrative: device evaluation by manufacturer: the aquabeam handpiece was not returned for investigation as it was disposed of at the user facility.A review of the device history record (dhr) ab2000-b / serial number: (b)(4) and aquabeam handpiece / lot number: 22c03392 was conducted, which confirmed that there were no non-conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system met all design and manufacturing specifications when released for distribution.The current user manual um0101-00 rev.F, aquabeam robotic system user manual, us, english was reviewed.Section 11.2.7 sterile: motorpack draping and docking with the aquabeam handpiece contains the note to, "verify the motorpack is securely engaged to the aquabeam handpiece by attempting to pull the aquabeam handpiece off of the motorpack.The aquabeam handpiece should remain connected to the motorpack if properly engaged." dock the motorpack to the aquabeam handpiece: o verify aquabeam handpiece nozzle position (i.E.The movement of the blue led) homes to the base of the aquabeamhandpiece (fully proximal).O apply sterile tape over the connection between aquabeam handpiece and motorpack to seal it.O keep the motorpack and the aquabeam handpiece assembly with the scope clamp assembly in a secure and sterile environment.The root cause of the reported event could not be established as the handpiece was not returned for investigation.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
900 island drive
suite 101
redwood city CA 94065 1494
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
900 island drive
suite 101
redwood city CA 94065 1494
Manufacturer Contact
doria esquivel
900 island drive
suite 101
redwood city, CA 94065-1494
MDR Report Key16218871
MDR Text Key309077353
Report Number3012977056-2023-00007
Device Sequence Number1
Product Code PZP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAB2000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/21/2022
Initial Date FDA Received01/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/07/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-