Model Number V60
Device Problem Charging Problem (2892)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2022
Event Type  malfunction  
Event Description
Philips received a complaint on the v60 ventilator indicating that the device was alarming.The device was reported to be outside of use, at the time of the reported problem.No patient harm reported.The customer informed the remote service engineer (rse) that the device was alarming low battery and the amber light on the display was not illuminated.The device was only running on the battery.The customer replaced the power cord and the device began to operate on ac power, but the battery still did not charge.Battery voltage was at11.39v and had a 2011 date code.The rse provided the customer replacement part information for the v60 battery.Good faith efforts (gfe) were made to confirm the part order and resolution and a gfe response received on 12jan2023 stated that the customer would order the battery.No further information was provided.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Manufacturer Narrative
This report is being submitted as a correction in response to corrective actions taken by the legal manufacturer.The manufacturer site number was incorrectly selected resulting in incorrect report numbers.A new mdr with the correct manufacturer site selected has been submitted under manufacturer report # 2518422-2023-17528.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key16219083
MDR Text Key308179484
Report Number2031642-2023-00261
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/28/2022
Initial Date FDA Received01/20/2023
Supplement Dates Manufacturer Received12/28/2022
Supplement Dates FDA Received07/31/2023
Date Device Manufactured09/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-