DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Component Incompatible (1108)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/14/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm), it was observed that the ac cord plug on the line cord assembly of the cardiosave intra-aortic balloon pump (iabp) was a non-compliant ac plug.There was no patient involvement.
|
|
Manufacturer Narrative
|
The getinge field service engineer (fse) that encountered the issue replaced the plug on line cord assembly (0012-00-1688-21), and completed the pm with full calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.This report is being submitted as the result of a retrospective review conducted in capa (b)(4).
|
|