• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. DISCOFIX®; ACCESSORIES, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. DISCOFIX®; ACCESSORIES, CATHETER Back to Search Results
Model Number 456020
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problem Low Blood Pressure/ Hypotension (1914)
Event Date 12/05/2022
Event Type  malfunction  
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number 400584376.The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event Description
As reported by the user facility: details of complaint (reported issue): after turning patient, i assessed femoral line and saw the port running vasopressors was backing up with blood and saw the stopcock had disconnected and the luer was completely broken off.2 stop cocks replaced with this patient.Patient became hypotensive an meds had to be modified.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Based on the data from the investigation we are unable to determine the root cause of the reported incident.The reported defect was unable to be confirmed.The actual defective device is a valuable tool in investigating the cause of this incident.Retained units were evaluated and passed the internal testing.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.Additionally, review of the batch history records was also performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOFIX®
Type of Device
ACCESSORIES, CATHETER
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key16223761
MDR Text Key308959966
Report Number2523676-2023-00010
Device Sequence Number1
Product Code KGZ
UDI-Device Identifier04046964996962
UDI-Public(01)04046964996962
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K760383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number456020
Device Catalogue Number456020
Device Lot Number21N06D9046
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/10/2023
Initial Date FDA Received01/23/2023
Supplement Dates Manufacturer Received01/10/2023
Supplement Dates FDA Received02/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-