• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-ELITE
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/07/2022
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to olympus for evaluation.The customer resolved the issue onsite by draining the air and replacing the water filter.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus that the water flow in the reprocessing basin decreased causing an increase in the reprocessing cycle time.It was the first cycle of the day.There were no reports of patient harm associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the water flow issue could not be determined.It is possible that the water flow decreased because the user did not read the instruction manual carefully and lacked knowledge of the replacement method of the water filter.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿8.4 replacing the water filter (maj-824 or maj-2318): caution: if air gets in the water filter housing, it may extend the process time.In case of an irregularity such as extension of the process time or lack of water supply, drain the air as described in ¿ draining air in the water filter housing¿ on page 402.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581-4003
5082077661
MDR Report Key16230780
MDR Text Key308367338
Report Number9610595-2023-01184
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170404047
UDI-Public04953170404047
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201920
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-ELITE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/07/2022
Initial Date FDA Received01/23/2023
Supplement Dates Manufacturer Received02/06/2023
Supplement Dates FDA Received02/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/30/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-