Model Number 0998-00-3023-53
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2023
Event Type  malfunction  
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) electrical test failure code 119, loud alarm.There was no patient involvement.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
Updated fields: b4, d9, g3, g6, g7, h2, h3, h4, h6, h10 a getinge field service engineer (fse) was dispatched to investigate the issue.In order to fix the issue, the fse replaced the main board pcb.The fse then performed a full preventive maintenance (pm).The unit passed the complete pm with full calibration, functional, and electrical safety tests to factory specifications.The unit was cleared for clinical use and returned to the customer.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16232843
MDR Text Key309013434
Report Number2249723-2023-00358
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 01/11/2023
Initial Date FDA Received01/24/2023
Supplement Dates Manufacturer Received07/20/2023
Supplement Dates FDA Received07/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/08/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-