ARROW INTERNATIONAL LLC ARROW EPIDURAL CATHETERIZATION KIT: 19 GA; ANESTHESIA CONDUCTION CATHETER Back to Search Results
Model Number IPN924964
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2023
Event Type  malfunction  
Event Description
Reported event: the catheter was occluded.There was no reported injury.A new catheter was inserted.
 
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Manufacturer Narrative
(b)(4).The customer returned one snaplock assembly, an epidural catheter, and lidstock.The returned components were received connected together.The returned components were visually examined with and without magnification.Visual examination of the returned snaplock assembly revealed the snaplock appears typical with no observed defects or anomalies.Visual examination of the returned catheter revealed the catheter appears used.Biological material can be seen between the inner coils.No other defects or anomalies were observed.A functional flow test was performed on the returned sample per amrq-000017 section 7.8; rev.10.The returned epidural catheter was inserted from the proximal end into the returned snaplock assembly until it bottomed out and the snaplock assembly was closed.The components were confirmed to be secured by tugging gently.The snaplock assembly was connected to the lab leak tester and the pressure was increased to 10 psi to establish flow.Water could be seen immediately exiting the distal end of the catheter.The flow rate was measured at 9.8ml/min, which is within the specification of 1ml/min minimum.No blockages were found.A device history record review was performed on the epidural catheter with no relevant findings.The reported complaint of the catheter being occluded could not be confirmed based on the sample received.A device history record review was performed on the epidural catheter with no evidence to suggest a manufacturing related issue.The returned components passed a functional flow test and met flow rate specifications.There were no functional issues found with the returned sample.No further action is required at this time.
 
Event Description
Reported event: the catheter was occluded.There was no reported injury.A new catheter was inserted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW EPIDURAL CATHETERIZATION KIT: 19 GA
Type of Device
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
312 commerce place
asheboro NC 27203
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194332672
MDR Report Key16234085
MDR Text Key308365870
Report Number1036844-2023-00001
Device Sequence Number1
Product Code BSO
UDI-Device Identifier10801902209501
UDI-Public10801902209501
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN924964
Device Catalogue NumberAK-05501
Device Lot Number13F22G0715
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/03/2023
Initial Date FDA Received01/24/2023
Supplement Dates Manufacturer Received02/20/2023
Supplement Dates FDA Received02/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NONE REPORTED; NONE REPORTED
-
-