MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION COUPLER; DEVICE, ANASTOMOTIC, MICROVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION COUPLER; DEVICE, ANASTOMOTIC, MICROVASCULAR Back to Search Results
Catalog Number 511100250010
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/13/2021
Event Type  malfunction  
Event Description
It was reported that a ring dislodged from a 2.5mm coupler.Additionally, during a surgical procedure "the couplers came out of their support when the vessels were positioned by the surgeon".The event occurred twice.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Initial reporter last name : (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the actual device was not available; however, five (5) companion samples were received for evaluation.The companion sample ring retention test was performed (force to pull the coupler ring from the jaw assembly) and the results were within the product specification range.The reported condition was not verified.The devices were conforming products.A device history review revealed no issues that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COUPLER
Type of Device
DEVICE, ANASTOMOTIC, MICROVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
SYNOVIS SURGICAL INNOVATIONS
2575 university ave. w
saint paul MN 55114
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16235708
MDR Text Key308751968
Report Number1416980-2023-00133
Device Sequence Number1
Product Code MVR
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K861985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511100250010
Device Lot NumberSP20J29-1485601
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/30/2022
Initial Date FDA Received01/24/2023
Supplement Dates Manufacturer Received02/15/2023
Supplement Dates FDA Received02/27/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-