Customer reports: after inserting a 10fr trocar chest tube into the infant's chest, c.Loughran was unable to remove the stylet/guidewire portion from the chest tube.Multiple attempts were made to remove the tube but it was not possible.The chest tube had to be removed and a new chest tube was inserted.Intervention was required for placement of the new chest due to the equipment malfunction of the original one.The patient was monitored throughout the procedure and no acute changes were noted by the providers and bedside staff.The chest tube was placed in a biohazard bag at the bedside, if needed for further investigation.Per additional information received on 1/24/23, the end of the device appeared to be tapered correctly and the wire insert also appeared to be embedded correctly.There was no patient injury and the medical intervention required was to pull and replace the chest tube.The patient has no issues regarding this incident.
|