• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC HEMOPRO POWER SUPPLY; UNIT, CAUTERY, THERMAL, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC HEMOPRO POWER SUPPLY; UNIT, CAUTERY, THERMAL, AC-POWERED Back to Search Results
Model Number C-VH-3010
Device Problem Failure to Deliver Energy (1211)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2022
Event Type  malfunction  
Manufacturer Narrative
Trackwise id (b)(6).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event Description
The hospital reported that the t.W.Power supply used for an endoscopic vein harvesting procedure stopped working.This was discovered at the beginning of a case prior to the device being used on the patient.Light and switch were on but not delivering energy.They did several troubleshooting things like switching power extension cables, etc.When they swapped power supplies the device began to work.No patient harm because no patient was involved.
 
Manufacturer Narrative
Trackwise # (b)(4).The reported device is an oem device.The certificate of conformance was reviewed for the serial # (b)(6).The vendor certifies that this device serial conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.H3 other text : device not returned.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOPRO POWER SUPPLY
Type of Device
UNIT, CAUTERY, THERMAL, AC-POWERED
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key16239932
MDR Text Key308362130
Report Number2242352-2023-00060
Device Sequence Number1
Product Code HQO
UDI-Device Identifier00607567700826
UDI-Public00607567700826
Combination Product (y/n)N
PMA/PMN Number
K052274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC-VH-3010
Device Catalogue NumberC-VH-3010
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/29/2022
Initial Date FDA Received01/25/2023
Supplement Dates Manufacturer Received04/17/2023
Supplement Dates FDA Received04/17/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/05/2017
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-