Model Number M3535A |
Device Problems
Device Alarm System (1012); Computer Software Problem (1112)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/15/2022 |
Event Type
malfunction
|
Event Description
|
It was reported to philips that the device displayed error codes in the status log.There was no reported patient impact or injury.
|
|
Manufacturer Narrative
|
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
|
|
Manufacturer Narrative
|
Philips received a complaint on the heartstart mrx indicating that there was an error message.There was reportedly no patient involvement.Remote support from the customer care solution center was received, during which a quote was provided for diagnostic service.It was determined that this was a malfunction of the processor pca for which the rse informed the customer that a replacement is necessary.The rse recommended replacing the processor pca to resolve the issue.The device remains at the customer site.No further action is warranted at this time.H3 other text : remote support provided.
|
|