Model Number AED PRO
Device Problem Incorrect Interpretation of Signal (1543)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2022
Event Type  malfunction  
Event Description
Complainant alleged that while attempting to defibrillate a 54-year-old female patient, the device issued a "shock advised" prompt for a heart rhythm they believe was non-shockable.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Manufacturer Narrative
Zoll medical corporation evaluated the device and the device successfully passed all testing.Review of the clinical files showed there was two analyses that were performed during the event, and shock was advised on one of the analysis.Zoll defibrillators function by taking a nine second ecg sample and dividing it into three, 3 seconds segments.Algorithm will make calculations based on ten waveform characteristics for each segment.For shock to advised a minimum of two of the three segments need to be identified as shockable.Review of the clinical files determined the aed pro device was functioning as it's designed and configured, within the limitations of the technology available.This investigation has been closed as device meets specification.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AED PRO
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key16241888
MDR Text Key308109198
Report Number1220908-2023-00134
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberAED PRO
Device Catalogue NumberAED PRO
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/04/2023
Initial Date FDA Received01/25/2023
Supplement Dates Manufacturer Received01/04/2023
Supplement Dates FDA Received03/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age54 YR
Patient SexFemale
-
-