Zoll medical corporation evaluated the device and the device successfully passed all testing.Review of the clinical files showed there was two analyses that were performed during the event, and shock was advised on one of the analysis.Zoll defibrillators function by taking a nine second ecg sample and dividing it into three, 3 seconds segments.Algorithm will make calculations based on ten waveform characteristics for each segment.For shock to advised a minimum of two of the three segments need to be identified as shockable.Review of the clinical files determined the aed pro device was functioning as it's designed and configured, within the limitations of the technology available.This investigation has been closed as device meets specification.No trend is associated with reports of this type.
|