• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ADVANTIO; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ADVANTIO; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number K064
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/15/2023
Event Type  Injury  
Manufacturer Narrative
The product has been received for analysis.This report will be updated upon completion of analysis.
 
Event Description
It was reported that this pacemaker was explanted and replaced due to entering safety mode.It was noted that during the replacement procedure, a temporary pacing system was used.No additional adverse patient effects were reported.
 
Event Description
It was reported that this pacemaker was explanted and replaced due to entering safety mode.It was noted that during the replacement procedure, a temporary pacing system was used.No additional adverse patient effects were reported.
 
Manufacturer Narrative
The product has been received for analysis.This report will be updated upon completion of analysis.-upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode due to system resets.It was confirmed that brady therapy remained available.The system resets caused the device to enter safety mode.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption related to telemetry attempts and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTIO
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16244890
MDR Text Key308144031
Report Number2124215-2023-03057
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N970003/S132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial,Followup
Report Date 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/06/2014
Device Model NumberK064
Device Catalogue NumberK064
Device Lot Number111721
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/15/2023
Initial Date FDA Received01/25/2023
Supplement Dates Manufacturer Received06/15/2023
Supplement Dates FDA Received07/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/08/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age63 YR
Patient SexFemale
-
-