U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD20
Device Problem Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2023
Event Type  malfunction  
Event Description
It has been reported to philips that the c-arm makes a grinding noise when moving, with bits of aluminum falling down.No harm was reported.We are conservatively reporting this event as the investigation is ongoing.A follow up report will be submitted when further information is received.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the system was not in clinical use when the issue was identified.The philips remote service engineer (rse) called and confirmed with customer that the c-arm makes grinding noise when moving.Rse was advised to clean the longitudinal rails and scheduled a fse visit.The philips field service engineer (fse) inspected the system onsite and confirmed the c arm was making noise and that some aluminum had fallen.To resolve the issue, the philips engineer replaced the z-rotation brakes.The system was performing as intended and was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16245926
MDR Text Key308188784
Report Number3003768277-2023-00555
Device Sequence Number1
Product Code IZI
UDI-Device Identifier00884838059054
UDI-Public00884838059054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD20
Device Catalogue Number722012
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/19/2023
Initial Date FDA Received01/26/2023
Supplement Dates Manufacturer Received12/04/2023
Supplement Dates FDA Received12/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/16/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-