The device log file was analyzed.At the time of the reported event a "motor stalled" entry was found indicating an issue with the ventilator motor.The device monitors the speed, rotation angle and other performance parameter of the motor continuously.Any speed fluctuations or other deviations may result in a divergence between expected piston position and the real hub that has been performed.The piston hub defines the applied tidal volume and thus, to prevent from potentially hazardous output and/or from damages to the ventilator unit, the system is designed to shut down automatic ventilation and to alert the user to this condition by means of a corresponding alarm.In case automatic ventilation is shut-down manual ventilation and monitoring functions remain available.The motor was replaced on-site and was provided for investigation.Different test scenarios were performed without provoking an error so that finally the root cause for the stalled motor could not be determined.Besides the issue that the patient could not be ventilated, a display failure was reported.Therefore, the control unit also comprising the display with backlight inverter was replaced and was analyzed at manufacturer site.Tests revealed a faulty backlight converter as root cause for the reported dark display.A non-functional display is obvious for the user.Normally, in case of a display failure automatic ventilation is continued.As for the current case automatic ventilation was stopped due to a stalled motor manual ventilation can be used to continue the procedure.The number of similar cases, related to the same root cause (backlight converter), is within the expected range of the respective risk assessment and thus accepted.As the root cause for the stalled motor could not be determined a corresponding assessment is not possible.Although both events (motor stalled and dark display) happened around the same time there were no hints found that the two events are related to each other.
|