SMITH & NEPHEW ORTHOPAEDICS AG POLARSTEM COLLAR STD. TI/HA 2; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED
|
Back to Search Results |
|
Model Number 75018402 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Hip Fracture (2349)
|
Event Date 04/25/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).The provided lot number is missing the last 4 characters for all hip prosthesis components.This complaint was opened by smith+nephew to document a patient complication identified through a review of the national joint registry from united kingdom that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
|
Event Description
|
It was reported that, after a right thr index surgery performed on (b)(6) 2020 to address an unspecified trauma injury, the patient sustained a periprosthetic fracture associated to the femoral component that made necessary a revision surgery on (b)(6) 2020.During this procedure, the femoral stem and head were explanted and replaced with a competitors femoral system.This information was provided by the national joint registry of the united kingdom, as part of a retrospective data collection of patients who underwent a primary thr surgery with a hip prosthesis construct that included a polarstem stem with an r3 acetabular cup and that required a revision surgery due to specific reasons.As such, no further information will be available.
|
|
Search Alerts/Recalls
|
|
|