• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED STERILIZATION PRODUCTS STERRAD® 100NX STERILIZER 1-DR; STERRAD® EQUIPMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED STERILIZATION PRODUCTS STERRAD® 100NX STERILIZER 1-DR; STERRAD® EQUIPMENT Back to Search Results
Model Number 10104
Device Problems Device Emits Odor (1425); Environmental Particulates (2930)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/07/2023
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone number: (b)(6).A field service engineer was dispatched to the customer site.The oil mist filter and o-ring were cleaned and reinstalled to resolve the mist/haze and odor/smells issues.Unit meets specifications and was returned to service.Investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the mist/haze and odor/smells issue, and system risk analysis (sra).The device history record (dhr) was reviewed and no issues relating to the failure mode were noted.The involved unit met manufacturer specifications at the time of release.Trending analysis of the mist/haze and odor/smells issues for the sterrad® 100nx unit was reviewed for the prior six months from the event date, and no significant trend was observed.Review of risk documentation shows the risk for exposure to toxic or corrosive material to be "low." no parts were replaced in resolving this issue, and therefore, no parts were available for further analysis.The assignable cause of the mist/haze and odor/smell is potentially due to the oil mist filter and o-ring.The asp field service engineer cleaned and reinstalled the parts and confirmed the sterrad® 100nx met functional specifications after service.The issues were resolved at the customer facility.Asp will continue to track and trend this issue.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by advanced sterilization products, or its employees that the report constitutes an admission that the product, advanced sterilization products, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Asp complaint ref #: (b)(4).
 
Event Description
A customer reported an event of a ¿mist¿ or haze emitting from the sterrad® 100nx sterilizer.There was no report of any injuries or human reactions.An asp field service engineer was dispatched to assess the unit onsite and identified an additional issue of odor/smell.This event is being reported as a malfunction report subsequent to a previous serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD® 100NX STERILIZER 1-DR
Type of Device
STERRAD® EQUIPMENT
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer (Section G)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer Contact
gabriela mclellan
33 technology drive
irvine, CA 92618
9495030264
MDR Report Key16253719
MDR Text Key308643991
Report Number2084725-2023-00011
Device Sequence Number1
Product Code MLR
UDI-Device Identifier10705037014491
UDI-Public10705037014491
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K071385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10104
Device Catalogue Number10104
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/07/2023
Initial Date FDA Received01/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/12/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-