MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE Back to Search Results
Catalog Number 787626
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/26/2022
Event Type  malfunction  
Event Description
It was reported that the shaft of the stent is severely bent and the device was not used.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Manufacturer Narrative
The reported event was unconfirmed because the reported failure could not be reproduced.No root cause could be found because the reported event was unconfirmed.A review of the device history record did not show any problems or condition that would have contributed to the reported issue.The instructions for use were found adequate and states the following: directions for use: 1.Method of use determine the proper stent length for the patient.This is generally calculated from the baseline pyelogram.Accurate measurements will optimize drainage efficiency and patient comfort.1 insertion of the guidewire : 1) remove the guidewire from the packaging, together with the guidewire holder.2) prior to removing the guidewire from the guidewire holder, inject sterile water through the port to activate the hydrophilic coating.3) remove the guidewire from the guidewire holder.Before using the guidewire, confirm the surface slides well when removing it from the holder.If you feel any resistance, do not force it, and inject the normal saline solution into the holder again, then try pulling it out once more.4) insert the guidewire either through the working channel of an endoscope or percutaneously, with the soft end of the guidewire first.5) advance the guidewire into the desired position until the tip is in the renal pelvis.Continuously confirm guidewire position either visually or under fluoroscopy.2.Precautions for use guidewire: (1)do not forcibly insert or remove the guidewire.It may injure patient or/and damage the device.(2)do not manipulate, advance and/or withdraw the guidewire through a metal cannula or needle; to do so may result in damage the guidewire.Avoid contact with devices with sharp edges (such as metal dilator).(3)avoid using of the device when resistance is encountered (due to the size of a catheter, stent and/or working-channel of endoscope) as this may cause wear the guidewire coating.(4)do not use organic medical solutions or oily contrast medium on this device.These solutions may damage the device or decrease the lubricity of the device.(5)the guidewire is treated with a hydrophilic coating.Do not insert a stent over the device with its surface insufficiently wet.Never use dry gauze.[hydrophilic polymer coating can be damaged, increasing resistance when trying to insert catheter, stent or endoscope.(6)if unusual resistance is met during manipulation of the guidewire, do not force to remove it.Carefully withdraw the guidewire as a unit.(7)do not use a retrieval device while the guidewire is in place; to do so may cause damage to the guidewire.(8)don¿t rub the guidewire with the edge of the holder.This could flake the hydrophilic coating.(9)avoid kinking, bending or twisting repeatedly the guidewire at acute bent site.It may lead to damage the guidewire.(10)never try to shape the guidewire.This could damage and break the cable core of the guidewire.(11)sufficient guidewire length must remain exposed to maintain a firm grip on the guidewire at all times.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the shaft of the stent is severely bent and the device was not used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE
Type of Device
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16254261
MDR Text Key308298937
Report Number1018233-2023-00344
Device Sequence Number1
Product Code FAD
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K043193
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number787626
Device Lot NumberMYGT5817
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/10/2023
Initial Date FDA Received01/27/2023
Supplement Dates Manufacturer Received03/09/2023
Supplement Dates FDA Received03/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-