Model Number 8884711253E
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2023
Event Type  malfunction  
Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
The customer reported that the enfit connector loosened.
 
Manufacturer Narrative
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.Two photos were provided for evaluation and the reported condition has been confirmed.A corrective and preventive action has been initiated to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOBBHOFF 12FR 43IN W STYLET EN
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key16259331
MDR Text Key308712773
Report Number9612030-2023-03539
Device Sequence Number1
Product Code KNT
UDI-Device Identifier10884521582927
UDI-Public10884521582927
Combination Product (y/n)N
Reporter Country CodeBR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884711253E
Device Catalogue Number8884711253E
Device Lot Number2105403564
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/10/2023
Initial Date FDA Received01/27/2023
Supplement Dates Manufacturer Received01/10/2023
Supplement Dates FDA Received02/20/2023
Date Device Manufactured03/09/2021
Is the Device Single Use? Yes
Type of Device Usage A
Patient Sequence Number1
-
-