This device is not distributed in us so that 510k# is blank.We checked the returned unit and confirmed that the ccd module blackout.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module.In addition, we confirmed that the ccd module fluid damage, the lg cable connector fluid damage, and the shield cover fluid damage; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|