(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
Customer returned pump for an alleged low bgs issues on (b)(6) 2022.The unit p-cap / test and reservoir locked in place properly.The pump passed the displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test, sleep current measurement, active current measurement, self test and displacement accuracy test.A water filled reservoir was used during testing.The pump was programmed with multiple boluses and monitored.All boluses delivered properly and were listed in the daily history.The pump was programmed with multiple basal profiles and monitored.All basal profiles delivered their indicated amounts and were verified in the daily and summary history screens.No delivery anomaly, bolus anomaly or basal anomaly noted during testing.Successfully downloaded history files and traces using thus software.The pump uploaded to carelink pro without any errors.Pump was cut open to perform visual inspection and found no evidence of physical or moisture damage on the pcba1, pcba2, force sensor, motor and battery tube assembly noted.The following were noted during visual inspection: minor scratches on lcd window, scratched case, cracked retainer and cracked keypad overlay.In summary, the pump passed functional testing.Unable to verify customer alleged for low bgs.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|