• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US EXPRESSEW III SUTURE PASSER W/ HOOK; SUTURE/NEEDLE PASSER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US EXPRESSEW III SUTURE PASSER W/ HOOK; SUTURE/NEEDLE PASSER, REUSABLE Back to Search Results
Model Number 214124
Device Problems Difficult to Open or Close (2921); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2022
Event Type  malfunction  
Event Description
It was reported by sales rep that during an unknown procedure on (b)(6) 2022, it was observed that the jaws on the expressew iii suture passer w/ hook device would not close properly.During in-house engineering evaluation, it was determined that the upper jaw on the device was loose; and could not be closed completely when the trigger was fully depressed.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).The expiration date is currently unavailable.Reporter is a j&j sales representative.Investigation summary: the product was returned to mitek for evaluation.Mitek then conducted visual inspection of device received.Upon visual inspection revealed wear marks on the device; also, the device does not show any structural anomalies.Upon reviewing the upper jaw, it was found that was loose.When the trigger was fully depressed, the jaw could not be closed completely.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection result, this complaint can be confirmed.The possible root cause for the loose jaw can be attributed to procedural variables, such handling of the device or product interaction during procedure, a bigger portion of tissue may have been grabbed and forced the jaw to close, also, since this device is reusable, the continuous use and sterilization process can cause metal fatigue leading to a non-closing jaw, however, this cannot be conclusively determined.As per ifu, it is important to inspect the device prior to use to ensure proper mechanical function and do not use if product is damaged.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.Device history lot: manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESSEW III SUTURE PASSER W/ HOOK
Type of Device
SUTURE/NEEDLE PASSER, REUSABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key16264447
MDR Text Key308959300
Report Number1221934-2023-00557
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10886705003655
UDI-Public10886705003655
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number214124
Device Catalogue Number214124
Device Lot Number66221
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/12/2023
Initial Date FDA Received01/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2022
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-