• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC 30K FSI-SLI-1000 INSERT,PKD; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC 30K FSI-SLI-1000 INSERT,PKD; SCALER, ULTRASONIC Back to Search Results
Catalog Number 81570
Device Problems Overheating of Device (1437); Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it is reported that a 30k fsi-sli-1000 insert,pkd allegedly runs hot during use, no matter the amount of water.No injury occurred.
 
Manufacturer Narrative
Returned qty.1 insert, 81570, 22255-00083084, 24, bul test method / gp005-wi02 inductance: gauge id# (b)(6) due: 11/30/2024 result: 3.16 meets spec does not meet spec n/a tip condition: - meets spec does not meet spec n/a stack condition meets spec does not meet spec n/a tested on: g130 gage id# (b)(6) due date: 09/30/2023 set pressure: 35 psi water flow: output rate: 35 psi - meets spec does not meet spec n/a spray pattern: - meets spec does not meet spec n/a water leak: hp sealing ring proximal ring distal ring seam leak n/a vibration: - meets spec does not meet spec n/a grip condition: meets spec does not meet spec n/a other visual observation: insert is good, no fault found.Also insert was tested on a digital thermometer i.D.#(b)(6) due date: 09/30/2023.The temperature was 84.0° f, no fault found.The specification states are not to exceed 118.4°f.Alleged event: confirmed - not confirmed root cause: no defect (aft0140) no defect.Also insert was tested on a digital thermometer i.D.#(b)(6) due date: 09/30/2023.The temperature was 84.0° f, no fault found.The specification states are not to exceed 118.4°f.Conclusion code: no failure found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30K FSI-SLI-1000 INSERT,PKD
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key16264606
MDR Text Key308857185
Report Number2424472-2023-00346
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K052334/K970
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 05/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number81570
Device Lot Number22255
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date01/27/2023
Initial Date Manufacturer Received 01/27/2023
Initial Date FDA Received01/30/2023
Supplement Dates Manufacturer Received01/27/2023
Supplement Dates FDA Received05/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-