Device evaluated by mfr.: the product was returned consisted of an angiojet spirolex catheter.The pump assembly, effluent/supply line, shaft, tip, and spike line were visually examined for damage or any irregularities.The catheter shaft showed multiple bends and kinks.Functional testing was competed per the instructions for use device preparation.The pump was inserted into the ultra drive unit console.The catheter primed, and the device functioned for a period of 90 seconds in the thrombectomy mode.The device pressure was within the normal range.No leaks were observed during functional testing.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|