Model Number ALLURA XPER FD10
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2023
Event Type  malfunction  
Event Description
It has been reported to philips that smoke was coming out of the review monitor.There was no report of harm.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the system was in clinical use when the issue occurred.The philips field service engineer (fse) inspected the system onsite and confirmed that smoke was coming out.A review of the system log file analyses provided does not confirm smoke coming out of the review monitor.Upon functional testing, fse found smoke was coming out of the review monitor.After replacing the new monitor with a new vesa adapter on the boom, the system was returned to good working order.These codes were updated based on investigation outcome.The evaluation method code grid was corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16268830
MDR Text Key308422507
Report Number3003768277-2023-00643
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD10
Device Catalogue Number722003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/17/2023
Initial Date FDA Received01/30/2023
Supplement Dates Manufacturer Received12/25/2023
Supplement Dates FDA Received12/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/03/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-