Model Number ALLURA XPER FD10
Device Problems Image Display Error/Artifact (1304); Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2023
Event Type  malfunction  
Event Description
It has been reported to philips that there is a strong stripe formation in live x-ray image.The device was in clinical use.It is unknown if the procedure was completed on the same device.Philips has started an investigation of the complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the system was used for planned treatment/non-emergency treatment when the issue occurred.The procedure was aborted.The philips field service engineer (fse) inspected the system onsite and confirmed that system had blurred images.Upon functional testing, the fse found that there was strong banding in the live xray image and the detector calibration failed.The philips engineer replaced the detector.After replacement, the system was returned to use in good working order.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16276083
MDR Text Key308528980
Report Number3003768277-2023-00655
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD10
Device Catalogue Number722003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/25/2023
Initial Date FDA Received02/01/2023
Supplement Dates Manufacturer Received06/12/2023
Supplement Dates FDA Received10/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/24/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-