• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number 2C8750
Device Problems Filling Problem (1233); Free or Unrestricted Flow (2945)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/06/2023
Event Type  malfunction  
Manufacturer Narrative
Age or date of birth: pediatric patient.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a clearlink system y-type blood/solution set was pulling fluid through but not blood.As a result, the patient received " a little more fluid" than expected.The roller clamp was as tight as it would go and the blood side was open.The issue was identified during patient infusion.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
B5: when using an iv pump, the saline was pulled down while the blood was infusing.H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device discarded; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16283287
MDR Text Key308629448
Report Number1416980-2023-00258
Device Sequence Number1
Product Code BRZ
UDI-Device Identifier00085412046341
UDI-Public(01)00085412046341
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8750
Device Lot NumberDR22F27093
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/06/2023
Initial Date FDA Received02/02/2023
Supplement Dates Manufacturer Received03/07/2023
Supplement Dates FDA Received03/08/2023
Date Device Manufactured06/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BLOOD; SALINE; UNSPECIFIED PUMP
-
-