Model Number TJF-Q190V
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2021
Event Type  malfunction  
Manufacturer Narrative
This mdr is being submitted as part of a retrospective review and remediation effort based on enhancements made to the company¿s mdr and complaint handling processes.Capas have been opened to manage the actions that are being taken to remediate this issue and ensure any required mdr reporting is completed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.During inspection and testing of the device, service could not confirm or duplicate the customer¿s reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, the following are the potential causes: deformation of the forceps channel.Breakage of the instrument.Foreign material in the forceps channel.The event can be detected/prevented by following the instructions for use: operation manual: 3.8 inspection of the endoscopic system.Inspection of the instrument channel and forceps elevator.Operation manual: important information ¿ please read before use; precautions.Operation manual: 3.6 inspection of ancillary equipment.Operation manual: 4.3 using endotherapy accessories.Reprocessing manual: 5.Reprocessing the endoscope (and related reprocessing accessories).Reprocessing manual: 7.Reprocessing endoscopes and accessories using an aer/wd.During the device evaluation, service found dents and scratches on the holder ring.The rubber adhesive on the insertion tube and distal end cover was cracked and peeling.Per the legal manufacturer, these other device issues identified by service have no potential to cause or contribute to death or serious injury if the malfunctions were to recur.Olympus will continue to monitor field performance for this device.
 
Event Description
The customer reported that a 10fr stent could not be passed through the instrument channel of the subject device.There was no patient or user injury reported.The device was returned to an olympus service center for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16284231
MDR Text Key308888899
Report Number9610595-2023-01673
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170452024
UDI-Public04953170452024
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K202661
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-Q190V
Device Catalogue NumberN5992941
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/28/2021
Initial Date FDA Received02/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-