• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH RADIAL HEAD W/ Ø22MM HD, Ø6.5MM STM,10(+3)MM HD HT,STER; PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH RADIAL HEAD W/ Ø22MM HD, Ø6.5MM STM,10(+3)MM HD HT,STER; PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER Back to Search Results
Catalog Number 09.405.263S
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Tissue Injury (4559)
Event Type  Injury  
Event Description
Device report from depuy synthes reports an event in switzerland as follows: it was reported that a year after implantation of the radial head replacement system (rhrs), the x-rays taken on an unknown date showed a light border (lysboarder) around the stem.Patient outcome is reported as happy and in good health.No further information provided.This report is for one (1) radial head w/ ø22mm hd, ø6.5mm stm,10(+3) mm hd ht,ster.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
No other medical intervention was required.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6: the device lot number is unknown, therefore a device history review could not be performed.If more information become available, the record will be re-assessed.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the device from the received image(s).Visual analysis of the photo revealed that there was no damage or defects with the rad head w/head ø22 shaft ø6.5 hh10 +3.There is not enough evidence to confirm the reported allegation based on the pictures provided.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the rad head w/head ø22 shaft ø6.5 hh10 +3 was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIAL HEAD W/ Ø22MM HD, Ø6.5MM STM,10(+3)MM HD HT,STER
Type of Device
PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wrights lane east
west chester, PA 19380
8472871282
MDR Report Key16287124
MDR Text Key308678683
Report Number8030965-2023-01342
Device Sequence Number1
Product Code KWI
UDI-Device Identifier07612334208019
UDI-Public(01)07612334208019
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K183618
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number09.405.263S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/24/2023
Initial Date FDA Received02/02/2023
Supplement Dates Manufacturer Received03/08/2023
04/03/2023
Supplement Dates FDA Received03/23/2023
04/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-