MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problems Mechanical Problem (1384); Device Difficult to Program or Calibrate (1496); Problem with Software Installation (3013)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/20/2022
Event Type  malfunction  
Event Description
It was reported that the programmer had a broken stylus and keyboard.It was noted the stylus was not functional.It was further reported that the programmer failed to update. there was no patient involvement.
 
Manufacturer Narrative
Analysis was able to partially confirm the customer comment, the software could not be remotely updated from the previous version however analysis identified that stylus and keyboard were fully functional and not broken.It was further identified that the upper and lower bullets were missing or worn.The cord bay,lid and upper and lower handles were broken.The keyboard cover plate was missing.All found defective parts were replaced and all other identified issues were resolved.The programmer then passed all final functional tests.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key16290926
MDR Text Key308727109
Report Number2182208-2023-00295
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2022
Initial Date Manufacturer Received 09/20/2022
Initial Date FDA Received02/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/28/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-